general knowledge of products
1、Summarize ans using
T&S vacuum steam autoclaves is small type one which according to Class B stardard. It can sterilize the solid, hollow, without package,single/double appliance,etc. which the big type do it.
It is very popular in dental clinique, internal medicine, E.N.T.department and laboratory etc.
|
2、The choose of dental line
In China,There are a lot of small dental clinique,the position is separate,it is difficulty to sterilize together.So the infections of dental are increasing. The sufferer needs the reliability of sterilize more and more. The sterilizing equipment become into the usual facility in dental clinique.
3、The importance of sterilize
In the dental therapy,80%-90% of the sufferer will bleeding during the processer, pathogen is inflectioned through the hand-piece and other appliance.The best way of sterilize recommended by WTO and CDC is high temperature vacuum steam sterilize.
4、The attentions of shoose steam autoclave
1)With the function of forvacuum,the best is adopt the tridimensional sterilizer.
The class B or S have the function of forvacuum,can expel the air mass from the chamber, insure sterilize the dental pieces drastically. The tridimensional sterilize could quick and exact controll, greatly reduce the damage.
2 )The veracity and evenness of the temperature
Veracity is the difference between chamber inner temperature and the setting temperature.(etc.121℃,134℃),The evenness is the difference around of the chamber.The sterilizer which don’t have good varacity and evenness will sterilize not cleanlily or damagethe hand-piece and other exact equipments.Our autoclave obtain temperature is 122 ℃± 1 ℃ or 135 ℃± 1 ℃。
3)Spare parts must be complete
The sterilizing equipments should inclding the ultrasonic cleaner,capper,sterilizing packing rolls and water distriller. Moreover choose the same brand.The parameter can easy matching,the after service is better.
5、The basis of choose the liter of Autoclave(referrence only)
1)If you sterilize the whole appliance, suggestion you adopt the 18L,23L. 2)You can make sure the liter according to the quantity of your cure machine.
|
quantity of cure machine |
1----3 pcs |
3----6pcs |
6----10pcs |
Above10 pcs |
|
Liter |
14L |
18L |
23L |
2pcs |
3)Purchase autoclave is a long investment, if your clinique will go on 3-5 years,you’d better buy a large one, the hospital which have 10pcs dental chait, It is better buy a large and small one.
6、The process of sterilizing
Clean outer → ultrasonic clean → cap package bag →high temperature sterilize →store
7、If the sterilize not successful,will infect the AIDS and other disease, use the correct way of sterilizing, not only couldn’t damage the dental appliance but also satisfied with the customer. |
Assemble and Use
Use the machine in the aeration enviroment, the crust mustn’t jam with anything. The space around the machine is 15cm, there are 20cm from the top. Please put the sterilizer on the lever table.
Before using the autoclave,you should study the simple guidelines first.
1、 first step:Join power line and drainpipe, add distilled water
Attention:
* The sterilizer’s heated element is high-power induction loop,when it is working,it will bring the higher inductance,so the electrical outlet must have secure grounding.
*Because the drainpipe is vent-pipe at the same time,so it must be buttoned.And must button up the waterspout.The gas that is vented has higher temperature,so it should be let in safe sewer or the vassel which is closed comparatively,lest penetrated in the room or scald somebody.The vent-pipe mustn’t be vent.
* The water must be distilled water.
If you collect waste water which be let by waste water tank,the drainpipe don’t dip in waste water,lest the waste water is absorbed in sterilizer by atmospheric pressure.
2、rn on the power,the sterilizer willraise the alarm,to tell you that it is short of water.Please press the add water key,the sterilizer will fill water in the water tank automatic.The water tank cubage is 2 lister, so you prepare the distilled water more than 2 lister.
3、ease press the open key,open the door,and put the goods on the plate.Please don’t sterilize without plate, if that will damage the dental pieces and also the manchine can’t work.
4、ess the door into the sterilizer for two seconds, the door will be locked automatically.
1)oose the sterilize program which you need, press the start key,then all will run automatically.
2) you have interest in and have enough time, you can observe the screen and will see the phase of sterilizing, state, correlative temperature,pressure,time and so on..
3)e in the process of sterilizing, if some data is error, the program will stop and tell you the error where or which was wrong.
4)anwhile after work, the printer(matching) will print the data of the sterilizing.
Attention:
Attention: Must use the clamp to take the plates out from the sterilizer to prevent scaled. You would better take the plates out after the temperature is get down.